FDA Adverse Event Malfunction Summary report: N

CHARGER?

MDR report key: 3843160 · Received June 2, 2014

Report

Report Number
2134265-2014-02995
Event Type
Malfunction
Date Received
June 2, 2014
Date of Event
May 6, 2014
Report Date
May 7, 2014
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
LIT
PMA / PMN Number
K112697
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR.: A VISUAL AND TACTILE EXAMINATION OF THE RETURNED DEVICE FOUND NO KINKS OR DAMAGE ALONG THE ENTIRE LENGTH OF THE DEVICE. POSITIVE PRESSURE WAS APPLIED AND LEAK WAS NOTED. A FURTHER MICROSCOPIC ANALYSIS CONFIRMED A LONGITUDINAL TEAR ALONG THE BALLOON LENGTH. NO OTHER DEFECTS WERE NOTED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(6).

Description of Event or Problem · 1

IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 7.0 X100, 135CM CHARGER BALLOON CATHETER WAS SELECTED TO DILATE THE LESION. DURING INFLATION THE BALLOON RUPTURED AT 11 ATMOSPHERES . THE DEVICE WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENTS STATUS WAS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321248 CHARGER? CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC - GALWAY H74939206071010 0016769899

Patients

Seq Age Sex Outcome Treatment
1