FDA Adverse Event Malfunction Summary report: N

DEXTRUS 4135

MDR report key: 3843143 · Received June 2, 2014

Report

Report Number
1028232-2014-01821
Event Type
Malfunction
Date Received
June 2, 2014
Date of Event
April 16, 2014
Report Date
May 21, 2014
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVN
PMA / PMN Number
P950037
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ASSOCIATED WITH THE MANUFACTURE OF THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT ATRIAL LEAD EXHIBITED LOSS OF CAPTURE DURING ROUTINE ATRIAL THRESHOLD TEST. IT WAS CONFIRMED THAT THE LEAD HAD MOVED FROM THE ORIGINAL IMPLANT POSITION. A REVISION PROCEDURE WAS PERFORMED WHEREIN THIS LEAD WAS SUCCESSFULLY REPOSITIONED AND HAS MAINTAINED THE REVISED POSITION TO DATE. THE FIELD REPRESENTATIVE INDICATED THAT AT THE TIME THAT LOC WAS DISCOVERED THERE WAS NO REPORT OF NOISE OR OVERSENSING; HOWEVER, A REDUCED P-WAVE WAS OBSERVED AT 0.4 MILLISECONDS. ASYSTOLE GREATER THAN 2 SECONDS WAS NOT DETECTED. THIS LEAD REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321932 DEXTRUS 4135 PACER LEAD NVN BIOTRONIK SE & CO. KG 358744

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization