FDA Adverse Event Malfunction Summary report: N

SMALL GRASPING RETRACTOR INSTRUMENT

MDR report key: 3843129 · Received June 2, 2014

Report

Report Number
2955842-2014-03357
Event Type
Malfunction
Date Received
June 2, 2014
Date of Event
April 29, 2014
Report Date
May 9, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ISI HAS RECEIVED THE INSTRUMENT INVOLVED WITH THIS COMPLAINT AND COMPLETED INVESTIGATIONS. FAILURE ANALYSIS INVESTIGATIONS CONFIRMED THE INITIAL REPORTER'S COMPLAINT. THE INSTRUMENT PITCH DOWN CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE BROKEN CABLE SEGMENT THAT CONTAINS THE CRIMP WAS MISSING FROM THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGES OR WEAR MARKS. NO OTHER DAMAGES WERE FOUND ON THE INSTRUMENT. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION OF BROKEN CABLE FOUND DURING FAILURE ANALYSIS, IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SIGMOID COLECTOMY PROCEDURE, THE SMALL GRASPING RETRACTOR BROKE. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. THERE WAS NO ALLEGATION OF ANY HARM, INJURY, OR ADVERSE OUTCOME TO A PATIENT. INTUITIVE SURGICAL, INC. (ISI) CONTACTED THE INITIAL REPORTER OF THIS COMPLAINT AND VERIFIED THAT THE CABLE WAS BROKEN ON THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
322479 SMALL GRASPING RETRACTOR INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420318-04 M10140124 718

Patients

Seq Age Sex Outcome Treatment
1