FDA Adverse Event Malfunction Summary report: N

ADULT CPB

MDR report key: 3843068 · Received March 14, 2014

Report

Report Number
2248146-2014-00027
Event Type
Malfunction
Date Received
March 14, 2014
Date of Event
February 1, 2014
Report Date
February 18, 2014
Manufacturer
DATASCOPE CORP.
Product Code
DWF
PMA / PMN Number
K090533
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WILL NOT BE RETURNED, THEREFORE NO EVAL CAN BE PERFORMED. IF THE DEVICE BECOMES AVAILABLE OR ADDITIONAL INFO IS RECEIVED, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THE CUSTOMER RECEIVED 20 BO-TOP 42802 PACKS AS A SUBSTITUTION TO THE REGULAR PACKS USED AT THIS FACILITY. IT WAS REPORTED THAT THE DESIGN OF THE PACK IS SUCH THAT THE PRE-BYPASS FILTER IS IN THE MIDDLE OF THE ARTERIAL-VENOUS LOOP, SEPARATING THE 3/8" ARTERIAL TUBING FROM THE 1/2" VENOUS TUBING. DESPITE CIRCULATING THE FLUID THROUGH THE CIRCUIT IN THE CUSTOMARY FASHION, SMALL BUBBLES PERSISTED ON THE VENOUS SIDE OF THE LOOP. PERFUSION STAFF TRIED A NUMBER OF DIFFERENT METHODS TO REMEDIATE, HOWEVER THE FINAL SOLUTION BECAME RE-ATTACHING THE ARTERIAL-VENOUS LOOP AFTER REMOVING THE PRE-BYPASS FILTER AND RE-CIRCULATING THE PRIME FLUID THROUGH THE CIRCUIT. IT WAS REPORTED THAT THE MALFUNCTION OCCURRED PRIOR TO USE ON A PT AND THERE WERE NO PT COMPLICATIONS AS A RESULT OF THIS FINDING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
154436 ADULT CPB CARDIOPULMONARY DEVICE DWF DATASCOPE CORP. BO-TOP 42802 15082-08

Patients

Seq Age Sex Outcome Treatment
1