HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2014-17622
- Event Type
- Malfunction
- Date Received
- June 2, 2014
- Date of Event
- May 22, 2014
- Report Date
- May 22, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. SHOULD THE DEVICE BE RETURNED OR ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A SYSTEM ERROR 2240 ALARM (AIR IN SET) OCCURRED ON THE HOMECHOICE (HC) DEVICE, DURING DWELL FIVE OF SIX. THE HOME PATIENT (HP) WAS CONNECTED AT THE TIME OF THE ALARM. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) HAD THE HP CLOSE ALL CLAMPS. THE TSR HAD THE HP CYCLE THE HC POWER OFF AND ON TO CLEAR THE ALARM. THERE WAS NOTHING UNUSUAL FOUND DURING THE TROUBLESHOOTING THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 322013 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HOMECHOICE |