FDA Adverse Event
Malfunction
Summary report: N
8800
MDR report key: 3842987
·
Received March 13, 2014
Report
- Report Number
- 1720753-2014-02286
- Event Type
- Malfunction
- Date Received
- March 13, 2014
- Date of Event
- February 14, 2014
- Report Date
- March 13, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE BATTERY PACK AND BATTERY CHARGER BOARD WERE EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A GENERATOR ERROR MESSAGE. THE SYSTEM AUTOMATICALLY SHUTS DOWN WHEN THIS OCCURS. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.