Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE GIB AND FFB WERE EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE FE REPORTED THAT THERE WAS A COMMUNICATION FAILURE ERROR MESSAGE. THIS ERROR IS LIKELY TO RESULT IN A SYSTEM LOCK UP, NO BOOT, SHUT DOWN DEPENDING ON WHEN THE LOSS OF COMMUNICATION OCCURED. THERE IS NO REPORT OF A PT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
151987 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800