FDA Adverse Event Malfunction Summary report: N

TERUMO ADVANCED PERFUSION SYSTEM 1

MDR report key: 3842938 · Received March 13, 2014

Report

Report Number
1828100-2014-00177
Event Type
Malfunction
Date Received
March 13, 2014
Date of Event
February 18, 2014
Report Date
February 18, 2014
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DTQ
PMA / PMN Number
K022947
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

EVALUATION IS IN PROGRESS, BUT NOT YET CONCLUDED.

Description of Event or Problem · 1

THE FIELD SERVICE REPRESENTATIVE (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, THE DRAW LATCH WAS BROKEN ON THE ULTRASONIC AIR SENSOR (UAS). THE UNIT STILL FUNCTIONED WITH THE BROKEN LATCH. SINCE THE EVENT OCCURRED DURING PREVENTATIVE MAINTENANCE, THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
152237 TERUMO ADVANCED PERFUSION SYSTEM 1 APS 1 (HEART LUNG CONSOLE) DTQ TERUMO CARDIOVASCULAR SYSTEMS CORP. 149673

Patients

Seq Age Sex Outcome Treatment
1