FDA Adverse Event Malfunction Summary report: N

PLUM A+ UK WITH 3PIN

MDR report key: 3842886 · Received March 10, 2014

Report

Report Number
9615050-2014-01824
Event Type
Malfunction
Date Received
March 10, 2014
Date of Event
January 1, 2014
Report Date
February 11, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K070398
Removal / Correction Number
FA212-08
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THE DEVICE DID NOT ALARM WHEN A DISTAL OCCLUSION WAS PRESENT. THIS WAS DUE TO A BROKEN DISTAL PRESSURE SENSOR PIN. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE DEVICE DISTAL PRESSURE SENSOR PIN WAS BROKEN. THERE WERE NO REPORTS OF ADVERSE PT EVENTS OR DELAYS IN CLINICAL USE. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
142005 PLUM A+ UK WITH 3PIN 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA PLUM A+ SOFTWARE MODULE, LIST #12097, SN (B)(4)