FDA Adverse Event
Malfunction
Summary report: N
PLUM A+ UK WITH 3PIN
MDR report key: 3842886
·
Received March 10, 2014
Report
- Report Number
- 9615050-2014-01824
- Event Type
- Malfunction
- Date Received
- March 10, 2014
- Date of Event
- January 1, 2014
- Report Date
- February 11, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K070398
- Removal / Correction Number
- FA212-08
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EI
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TESTING AND INVESTIGATION FOUND THE DEVICE DID NOT ALARM WHEN A DISTAL OCCLUSION WAS PRESENT. THIS WAS DUE TO A BROKEN DISTAL PRESSURE SENSOR PIN. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE DEVICE DISTAL PRESSURE SENSOR PIN WAS BROKEN. THERE WERE NO REPORTS OF ADVERSE PT EVENTS OR DELAYS IN CLINICAL USE. NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 142005 | PLUM A+ UK WITH 3PIN | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | PLUM A+ SOFTWARE MODULE, LIST #12097, SN (B)(4) |