SYSTEM 7 V-NOTCH SAGITTAL SAW
Report
- Report Number
- 0001811755-2014-02008
- Event Type
- Malfunction
- Date Received
- June 2, 2014
- Date of Event
- May 23, 2014
- Report Date
- May 23, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HWE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED. FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.
THE REPORTED OVERHEATING OF THE DEVICE WAS DUPLICATED BY A MANUFACTURER REPAIR TECHNICIAN THROUGH FUNCTIONAL EVALUATION. UPON DISASSEMBLY, IT WAS FOUND THAT THE DRIVE LINK WAS LOOSE, WHICH CAN LEAD TO THE REPORTED EVENT AND CAN BE CAUSED BY A MATERIAL FATIGUE ISSUE OR IMPACT.
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY THE DEVICE WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING BACK-UP EQUIPMENT. NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY, THE DEVICE WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING BACK-UP EQUIPMENT. NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.