DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DOES NOT RECOGNIZE BATTERY PACK) WAS CONFIRMED. UPON EVALUATION, THERE WAS CONTAMINATION ON THE BATTERY BOARD. THE CAUSE OF THE INABILITY TO RECOGNIZE A BATTERY PACK IS THE CONTAMINATION. THE ROOT CAUSE OF THE CONTAMINATION COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY LIQUID INGRESS OF AN UNK CONTAMINANT. NO ADVERSE EVENT RESULTED FROM THE CONTAMINATED CHARGER/MODEM. THE PT RECEIVED A REPLACEMENT CHARGER/MODEM.

Description of Event or Problem · 1

A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE CHARGER/MODEM WOULD NOT RECOGNIZE A BATTERY PACK. THE PT WAS ISSUED A REPLACEMENT BATTERY PACK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
109410 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 69 YR