FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3842560
·
Received February 21, 2014
Report
- Report Number
- 3008642652-2014-00504
- Event Type
- Malfunction
- Date Received
- February 21, 2014
- Date of Event
- January 31, 2014
- Report Date
- February 12, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DOES NOT RECOGNIZE BATTERY PACK) WAS CONFIRMED. UPON EVALUATION, THERE WAS CONTAMINATION ON THE BATTERY BOARD. THE CAUSE OF THE INABILITY TO RECOGNIZE A BATTERY PACK IS THE CONTAMINATION. THE ROOT CAUSE OF THE CONTAMINATION COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY LIQUID INGRESS OF AN UNK CONTAMINANT. NO ADVERSE EVENT RESULTED FROM THE CONTAMINATED CHARGER/MODEM. THE PT RECEIVED A REPLACEMENT CHARGER/MODEM.
Description of Event or Problem · 1
A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE CHARGER/MODEM WOULD NOT RECOGNIZE A BATTERY PACK. THE PT WAS ISSUED A REPLACEMENT BATTERY PACK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 109410 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR |