FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3842559 · Received February 21, 2014

Report

Report Number
3008642652-2014-00503
Event Type
Malfunction
Date Received
February 21, 2014
Date of Event
January 21, 2014
Report Date
February 12, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY CHARGER SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WON'T POWER ON) HAS BEEN CONFIRMED. AS RECEIVED, THE CHARGER/MODEM WOULD NOT POWER ON. UPON EVALUATION, THE POWER SUPPLY UNIT WAS DEFECTIVE. THE CAUSE OF THE INABILITY TO POWER ON IS THE DEFECTIVE POWER SUPPLY UNIT. THE ROOT CAUSES OF THE DEFECTIVE POWER SUPPLY UNIT CANNOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE POWER SUPPLY. THE PT WAS ISSUED A REPLACEMENT BATTERY CHARGER.

Description of Event or Problem · 1

A (B)(6) FEMALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE BATTERY CHARGER WOULD NOT POWER ON. THE PT WAS ISSUED A REPLACEMENT CHARGER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
109862 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA