FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3842559
·
Received February 21, 2014
Report
- Report Number
- 3008642652-2014-00503
- Event Type
- Malfunction
- Date Received
- February 21, 2014
- Date of Event
- January 21, 2014
- Report Date
- February 12, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY CHARGER SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WON'T POWER ON) HAS BEEN CONFIRMED. AS RECEIVED, THE CHARGER/MODEM WOULD NOT POWER ON. UPON EVALUATION, THE POWER SUPPLY UNIT WAS DEFECTIVE. THE CAUSE OF THE INABILITY TO POWER ON IS THE DEFECTIVE POWER SUPPLY UNIT. THE ROOT CAUSES OF THE DEFECTIVE POWER SUPPLY UNIT CANNOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE POWER SUPPLY. THE PT WAS ISSUED A REPLACEMENT BATTERY CHARGER.
Description of Event or Problem · 1
A (B)(6) FEMALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE BATTERY CHARGER WOULD NOT POWER ON. THE PT WAS ISSUED A REPLACEMENT CHARGER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 109862 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |