FDA Adverse Event Malfunction Summary report: N

MICRO OSCILLATING SAW

MDR report key: 3842536 · Received June 2, 2014

Report

Report Number
0001811755-2014-01995
Event Type
Malfunction
Date Received
June 2, 2014
Date of Event
May 6, 2014
Report Date
May 6, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.

Additional Manufacturer Narrative · 1

THE TECHNICIAN CONFIRMED THE REPORTED EVENT AND NOTED THAT THE MOTOR WAS CORRODED.

Description of Event or Problem · 1

THE MICRO OSCILLATING SAW WAS RETURNED TO STRYKER INSTRUMENTS FOR SERVICE. DURING FUNCTIONAL TESTING BY A SERVICE TECHNICIAN, IT WAS FOUND THAT THE DEVICE CAUSED A BIAS CURRENT MESSAGE TO BE DISPLAYED ON THE CONSOLE, SIGNALING A CONDITION OCCURRED IN WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

Description of Event or Problem · 1

THE MICRO OSCILLATING SAW WAS RETURNED TO STRYKER INSTRUMENTS FOR SERVICE. DURING FUNCTIONAL TESTING BY A SERVICE TECHNICIAN, IT WAS FOUND THAT THE DEVICE CAUSED A BIAS CURRENT MESSAGE TO BE DISPLAYED ON THE CONSOLE, SIGNALING A CONDITION OCCURRED IN WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321094 MICRO OSCILLATING SAW DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE ERL STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1