FDA Adverse Event Malfunction Summary report: N

SP2 TIBIAL RADEL IMPACTOR

MDR report key: 3842468 · Received June 2, 2014

Report

Report Number
1818910-2014-19851
Event Type
Malfunction
Date Received
June 2, 2014
Date of Event
May 13, 2014
Report Date
May 13, 2014
Manufacturer
DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDICS ? REG.
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: EXAMINATION OF THE SUBMITTED IMPACTOR CONFIRMED THE FRACTURE INITIATION ALONG THE SLOT THAT CONNECTS WITH THE UNIVERSAL HANDLE. THE ROOT CAUSE IS ATTRIBUTED TO MISUSE THROUGH MISALIGNMENT. THE FLAT OF THE IMPACTOR EXHIBITS GOUGING INDICATING THE IMPACTOR WAS NOT PROPERLY ALIGNED PRIOR TO IMPACTION. NO EVIDENCE WAS FOUND INDICATING PRODUCT ERROR WAS A CONTRIBUTING FACTOR. BASED ON THE PERFORMED INVESTIGATION, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. MONITOR THROUGH TREND ANALYSIS. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

HP TIBIAL IMPACTOR TIP BROKE AT THE TOP WHERE IT CONNECTS TO THE UNIVERSAL HANDLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321016 SP2 TIBIAL RADEL IMPACTOR KNEE INSTRUMENT/TRIAL LXH DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDICS ? REG. J0608

Patients

Seq Age Sex Outcome Treatment
1