FDA Adverse Event Malfunction Summary report: N

PROGRASP FORCEPS INSTRUMENT

MDR report key: 3842294 · Received June 1, 2014

Report

Report Number
2955842-2014-03347
Event Type
Malfunction
Date Received
June 1, 2014
Date of Event
May 7, 2014
Report Date
May 7, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION OBSERVED THAT THE PITCH UP CABLE WAS FOUND TO BE FRAYED AT THE DISTAL CLEVIS HUB. A SMALL FRAYED STRANDS STUCK OUT AT THE WRIST. OTHER CABLES AT WRIST WERE UNDAMAGED AND CLEVIS SHOWED NO DAMAGE. AN ADDITIONAL OBSERVATION NOT REPORTED BY SITE WAS THAT THE INSTRUMENT'S MAIN TUBE EXHIBITED SCRATCH MARKS/ABRASIONS. THE DISTAL END OF MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE .265 - .149 IN LENGTH AND NOT ALIGNED WITH THE TUBE AXIS. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE DAMAGE TO THE INSTRUMENT'S PITCH UP AND/OR MAIN TUBE FOUND DURING FAILURE ANALYSIS INVESTIGATION COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING REPROCESSING OF THE PROGRASP FORCEPS INSTRUMENT, IT WAS NOTED THAT THE CABLE ON THE INSTRUMENT WAS FRAYED. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT. THERE WAS NO ALLEGATION OF ANY HARM, INJURY, OR ADVERSE OUTCOME TO A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
320701 PROGRASP FORCEPS INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420093-10 M10131024 179

Patients

Seq Age Sex Outcome Treatment
1