ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-15395
- Event Type
- Malfunction
- Date Received
- May 31, 2014
- Report Date
- May 27, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
(B)(4). DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: POWER ON RESETS OBSERVED IN THE BLACK BOX. BATTERY COMPARTMENT IS CRACKED BELOW THE BUMPER GRIP. NO DAMAGE FOUND TO RETURNED BATTERY CAP. THE RETURNED BATTERY CAP CONTACT MEASUREMENTS ARE IN SPECIFICATIONS. PUMP WAS EXERCISED FOR 24HR DURATION PERIOD USING THE RETURNED BATTERY CAP; NO POWER INTERRUPTIONS WERE DUPLICATED DURING TESTING WITH THE RETURNED BATTERY CAP. REMOVAL OF THE PUMP COVER; SHOWED NO EVIDENCE OF INTERNAL MOISTURE. NO DAMAGE TO THE POWER CIRCUIT WAS OBSERVED. THE COMPLAINT WAS CONFIRMED IN PUMP HISTORY BUT NOT DUPLICATED DURING TESTING. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 320367 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |