ANIMAS VIBE
Report
- Report Number
- 2531779-2014-15383
- Event Type
- Malfunction
- Date Received
- May 31, 2014
- Report Date
- May 21, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1: DATE OF SUBMISSION 09/10/2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: LOSS OF PRIME WARNINGS WITH ZERO FORCE EVENTS WERE OBSERVED AT THE END OF THE BLACK BOX HISTORY. THE PUMP REWIND, LOAD CARTRIDGE, AND PRIME STEPS WERE PERFORMED SUCCESSFULLY WITH NO ALARMS. A FORCE SENSOR CALIBRATION TEST CONFIRMED THAT THE SENSOR WAS DETECTING CORRECT FORCE AT 5 POUNDS. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO LOSS OF PRIMES OCCURRING. THE PUMP WAS OPENED AND NO DAMAGE WAS FOUND TO THE FORCE SENSOR CIRCUIT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. THE REPORTER STATED THAT FREQUENT LOSS OF PRIME EVENTS HAD OCCURRED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 320467 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |