FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 3842206 · Received May 31, 2014

Report

Report Number
2531779-2014-15364
Event Type
Malfunction
Date Received
May 31, 2014
Report Date
May 26, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 08/05/2014-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/23/2014 WITH THE FOLLOWING FINDINGS:A REVIEW OF THE PUMP¿S BLACK BOX SHOWED EVIDENCE OF SUDDEN VOLTAGE DROPS ON 05/07/2014. MULTIPLE BATTERY RELATED ALARMS WERE OBSERVED IN THE ALARM HISTORY. DURING TESTING THE PUMP WAS ABLE TO POWER ON AND PRIME CORRECTLY. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO ALARMS. THE CURRENT DRAWS WERE FOUND TO BE ABOVE SPECIFICATIONS. AN INTERMITTENT CONNECTION WAS FOUND ON THE PRINTED CIRCUIT BOARD. THE CONNECTION WAS RE-ESTABLISHED AND THE CURRENT DRAWS WERE NORMAL. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE) ISSUE. IT WAS REPORTED THAT THE BATTERY LIFE WAS SHORTER THAN EXPECTED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
320535 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1