ANIMAS VIBE
Report
- Report Number
- 2531779-2014-15364
- Event Type
- Malfunction
- Date Received
- May 31, 2014
- Report Date
- May 26, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP # 1 DATE OF SUBMISSION 08/05/2014-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/23/2014 WITH THE FOLLOWING FINDINGS:A REVIEW OF THE PUMP¿S BLACK BOX SHOWED EVIDENCE OF SUDDEN VOLTAGE DROPS ON 05/07/2014. MULTIPLE BATTERY RELATED ALARMS WERE OBSERVED IN THE ALARM HISTORY. DURING TESTING THE PUMP WAS ABLE TO POWER ON AND PRIME CORRECTLY. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO ALARMS. THE CURRENT DRAWS WERE FOUND TO BE ABOVE SPECIFICATIONS. AN INTERMITTENT CONNECTION WAS FOUND ON THE PRINTED CIRCUIT BOARD. THE CONNECTION WAS RE-ESTABLISHED AND THE CURRENT DRAWS WERE NORMAL. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE) ISSUE. IT WAS REPORTED THAT THE BATTERY LIFE WAS SHORTER THAN EXPECTED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 320535 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |