ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-15355
- Event Type
- Malfunction
- Date Received
- May 31, 2014
- Report Date
- May 27, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP # 1 DATE OF SUBMISSION 06/24/2014-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/20/2014 WITH THE FOLLOWING RESULTS:A REVIEW OF THE USER¿S SETTINGS SHOWED THAT THE ¿BASAL PROGRAM N1¿ WAS SET TO ¿0.000U/HOUR¿ AT 1:30PM, THE BASAL HISTORY REFLECTED 0.0U/HOUR EVERY DAY AT 1:32PM SINCE 05/07/2014. DURING TESTING, THE PUMP AND THE BASAL HISTORY REFLECTED 24U AFTER 24 HOURS ON A 1U/ HOUR BASAL PROGRAM DURATION TEST. THE PUMP WAS FOUND TO OPERATE WITHIN SPECIFICATIONS.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A HISTORY/SETTINGS ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 320574 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 10 YR |