FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3842189 · Received May 31, 2014

Report

Report Number
2531779-2014-15355
Event Type
Malfunction
Date Received
May 31, 2014
Report Date
May 27, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 06/24/2014-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/20/2014 WITH THE FOLLOWING RESULTS:A REVIEW OF THE USER¿S SETTINGS SHOWED THAT THE ¿BASAL PROGRAM N1¿ WAS SET TO ¿0.000U/HOUR¿ AT 1:30PM, THE BASAL HISTORY REFLECTED 0.0U/HOUR EVERY DAY AT 1:32PM SINCE 05/07/2014. DURING TESTING, THE PUMP AND THE BASAL HISTORY REFLECTED 24U AFTER 24 HOURS ON A 1U/ HOUR BASAL PROGRAM DURATION TEST. THE PUMP WAS FOUND TO OPERATE WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A HISTORY/SETTINGS ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
320574 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 10 YR