FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3842173 · Received May 31, 2014

Report

Report Number
2531779-2014-15341
Event Type
Malfunction
Date Received
May 31, 2014
Report Date
May 23, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 07/16/2014 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION REVEALED THAT THERE WAS EVIDENCE OF MOISTURE DAMAGE UNDER THE DISPLAY LENS AND THAT THE AUDIO BOLUS BUTTON COVER WAS MISSING/DETACHED. A LEAK TEST WAS PERFORMED AND A LEAK WAS FOUND AT THE MISSING AUDIO BOLUS BUTTON COVER. THE PUMP CASE WAS REMOVED AND EVIDENCE OF MOISTURE DAMAGE WAS OBSERVED. THE COMPLAINT OF THE UNDER RESPONSIVE KEYPAD BUTTONS WAS NOT ABLE TO BE DUPLICATED DURING INVESTIGATION. ALL KEYPAD BUTTONS WERE UNRESPONSIVE TO BUTTON PRESSES. THE KEYPAD COVER WAS REMOVED AND NO EVIDENCE OF CONTAMINATION WAS FOUND UNDER ANY BUTTON CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES W/MOISTURE) ISSUE. THE REPORTER ALLEGED THAT ALL KEYPAD BUTTONS WERE UNDER RESPONSIVE. IT WAS NOTED THAT THERE WAS MOISTURE/CORROSION IN THE BATTERY COMPARTMENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
320455 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 18 YR