FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3842154 · Received May 31, 2014

Report

Report Number
2531779-2014-15312
Event Type
Malfunction
Date Received
May 31, 2014
Report Date
May 26, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 07/01/2014-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/26/2014 WITH THE FOLLOWING FINDINGS:DURING A VISUAL INSPECTION OF THE PUMP, MOISTURE WAS VISIBLE BEHIND THE DISPLAY LENS. NO DAMAGE WAS OBSERVED TO THE PUMP BATTERY COMPARTMENT, AND THE BATTERY CAP SECURED PROPERLY. THE PUMP CASING WAS OBSERVED TO BE CRACKED IN THE UPPER RIGHT CORNER NEAR THE DISPLAY. DURING TESTING, THE PUMP POWERED ON WITH AUDIBLE TONES, NO VIBRATION, AND A BLANK DISPLAY. A LEAK TEST WAS PERFORMED AND CONFIRMED A LEAK AT THE CRACK IN THE PUMP CASING. THE PUMP CASE WAS REMOVED AND EVIDENCE OF MOISTURE INGRESS WAS FOUND THROUGHOUT THE PUMP INTERIOR.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (MOISTURE INGRESS) ISSUE. IT WAS REPORTED THAT THE PUMP LOST POWER. MOISTURE WAS ALLEGEDLY VISIBLE BEHIND THE PUMP DISPLAY LENS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
320520 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 9 YR