ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-15305
- Event Type
- Malfunction
- Date Received
- May 31, 2014
- Report Date
- May 26, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PATIENT
Narratives
(B)(4)-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP BLACK BOX DATA REVEALED A PUMP REBOOT. DURING A VISUAL INSPECTION OF THE PUMP, THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED. THE RETURNED BATTERY CAP THREADS WERE ALSO FOUND TO BE STRIPPED. DURING TESTING, THE BATTERY CAP DID NOT SECURE PROPERLY TO THE PUMP AND A POWER LOSS WAS DUPLICATED. A TEST BATTERY CAP SECURED PROPERLY TO THE PUMP. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER INTERRUPTIONS. THE PUMP CASE WAS REMOVED, AND NO EVIDENCE OF MOISTURE INGRESS OR DAMAGE TO THE POWER CIRCUIT WAS FOUND.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (DAMAGE) ISSUE. IT WAS REPORTED THAT THE PUMP INTERMITTENTLY LOST POWER. ALLEGEDLY, THE BATTERY COMPARTMENT WAS CRACKED, AND THE YELLOW O-RING OF THE BATTERY CAP WAS VISIBLE WHILE THE CAP WAS ATTACHED TO THE PUMP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 320306 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR |