GREENLIGHT MOXY FIBER OPTIC
Report
- Report Number
- 2937094-2014-00429
- Event Type
- Malfunction
- Date Received
- May 31, 2014
- Date of Event
- April 22, 2014
- Report Date
- May 19, 2014
- Manufacturer
- AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
- Product Code
- GEX
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER (B)(4): THE REPORT OF "CAP DETACHED" COULD NOT BE CONFIRMED; HOWEVER, THE FIBER EXHIBITED A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE AT THE BEVEL EDGE; THE GLASS/METAL CAPS REMAINED INTACT AND ATTACHED; THE FIBER PROXIMAL TO FRACTURE ROTATED INDEPENDENTLY OF METAL CAP; THE GLASS CAP EXHIBITED SEVERE DEVITRIFICATION AT THE OUTPUT WINDOW; THE METAL CAP EXHIBITED SEVERE BURNT ON DETRITUS; THE OUTER FLOW TUBING EXHIBITED SEVERE CONTAMINATION, LIKELY BIOLOGIC. BASED ON THE ANALYSIS ABOVE, THE POTENTIAL FOR FORWARD FIRING MAY ALSO EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH LIMITING THE PERFORMANCE OF THE FIBER.
IT WAS REPORTED THAT DURING A PROSTATE PROCEDURE, THE FIBER CAP DETACHED AT 216,000 JOULES AND 30 MINUTES OF USAGE. THE PROCEDURE WAS COMPLETED USING A SECOND FIBER. ¿NO DAMAGE TO THE PATIENT¿ WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 320575 | GREENLIGHT MOXY FIBER OPTIC | POWERED LASER SURGICAL INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) | 0010-2400 | 347A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |