STERRAD NX STERILIZER
Report
- Report Number
- 2084725-2014-00199
- Event Type
- Malfunction
- Date Received
- May 30, 2014
- Date of Event
- May 19, 2014
- Report Date
- May 19, 2014
- Manufacturer
- ADVANCED STERILIZATION PRODUCTS
- Product Code
- MLR
- PMA / PMN Number
- K042116
- Removal / Correction Number
- Z-0744-0746-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER WAS DISPATCHED TO CUSTOMER SITE. ODOR/SMELL WAS CONFIRMED. THE VACUUM PUMP OIL, CATALYTIC DECOMP FILTER AND OIL MIST FILTER WERE REPLACED. UNIT MEETS SPECIFICATIONS AND WAS RETURNED TO SERVICE.
MANUFACTURER DATE: 09/30/2008. CONCLUSION: BASED ON CAPA INVESTIGATION AND ASTM TESTING, OIL DEGRADATION CAN CAUSE ODOR/SMELLS FOR STERRAD SYSTEMS. H3: ASP INVESTIGATION SUMMARY: THE INVESTIGATION INCLUDED A REVIEW OF THE DEVICE HISTORY RECORD (DHR), SERVICE HISTORY, TRENDING OF THE PRODUCT MALFUNCTION CODE, FAILURE MODE AND EFFECTS ANALYSIS (FMEA), AND SYSTEM HAZARD AND USER MISUSE ANALYSIS (SHUMA), AND CORRECTIVE AND PREVENTATIVE ACTION (CAPA). ¿ THE DHR WAS REVIEWED. THE UNIT MET MANUFACTURING SPECIFICATIONS AT THE TIME OF RELEASE AND THERE WERE NO ISSUES RELATED TO THIS FAILURE MODE NOTED. ¿ THE SERVICE HISTORY FOR THIS UNIT FOR THE PAST SIX MONTHS (12/01/2013 TO 05/30/2014) DID NOT OBSERVE ANY SIGNIFICANT TREND. ¿ THE TREND FOR THE PRODUCT MALFUNCTION CODE OF ODOR/SMELLS WAS COMPLETED FROM MAY 2013 THROUGH APRIL 2014 AND REVEALED A SIGNIFICANT TREND WHICH IS ADDRESSED IN CAPA. ¿ THE FMEA REVEALED THE RISK PRIORITY NUMBER (RPN) SCORES ARE BELOW 100 AND IS CONSIDERED ACCEPTABLE. ¿ THE SHUMA INDICATES THE RISK IS BROADLY ACCEPTABLE FOR MODERATE LEVEL INJURY AND AS LOW AS REASONABLY PRACTICABLE FOR MILD (LIMITED) EXPOSURE TO ODOR OR ODORANTS THAT ARE SELF-MANAGED. ¿ CAPA INVESTIGATION AS WELL AS THE ASTM TESTING PERFORMED INDICATED OIL DEGRADATION CAN CAUSE ODOR/SMELLS FOR STERRAD® SYSTEMS. NO PARTS WERE RETURNED FOR FURTHER EVALUATION THEREFORE, THE CAUSE WAS NOT CONFIRMED. THE REPORTED ISSUE WAS RESOLVED AT THE CUSTOMER FACILITY. THE ISSUE WILL BE TRACKED AND TRENDED. NO FURTHER INVESTIGATION IS NECESSARY AT THIS TIME.
AN CUSTOMER REPORTED AN EVENT OF AN ODOR/SMELL EMITTING FROM THE STERRAD® NX STERILIZER. THERE WAS NO REPORT OF ANY INJURIES OR HUMAN REACTIONS. A FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY EVENT DATED (B)(6) 2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 318953 | STERRAD NX STERILIZER | STERRAD EQUIPMENT (MLR) | MLR | ADVANCED STERILIZATION PRODUCTS | NA | 081252 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |