FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3841773 · Received May 30, 2014

Report

Report Number
2032227-2014-02700
Event Type
Malfunction
Date Received
May 30, 2014
Date of Event
May 9, 2014
Report Date
May 12, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT INSULIN PUMP IS NOT DELIVERING THE CORRECT AMOUNT OF INSULIN. DURING THE TROUBLESHOOTING, NO ALARMS WERE FOUND, BUT THE CUSTOMER IS STILL EXPERIENCING HIGH BLOOD GLUCOSE LEVELS. THE BLOOD GLUCOSE READING WAS 430 MG/DL. CUSTOMER EXPERIENCED EXCESSIVE THIRST AND FATIGUE. CUSTOMER TREATED WITH MANUAL INJECTION. NO FURTHER INFORMATION WAS REPORTED.