FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3841662 · Received May 30, 2014

Report

Report Number
2032227-2014-02613
Event Type
Injury
Date Received
May 30, 2014
Date of Event
May 4, 2014
Report Date
May 4, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WV
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. HOWEVER, THE INSULIN PUMP WAS RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNER, MINOR SCRATCHED LCD WINDOW AND CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

THE CUSTOMER REPORTS SHE WAS HOSPITALIZED WITH HIGH BLOOD GLUCOSE OF 583 MG/DL. AT THE TIME OF THE REPORT, THE CUSTOMER'S BLOOD GLUCOSE HAD LOWERED TO 268 MG/DL. THE CUSTOMER STATES HER INSULIN PUMP DID NOT DELIVER INSULIN ALL DAY. ASSISTANCE WAS PROVIDED IN CLEARING A MOTOR ERROR ALARM. THE INSULIN PUMP WILL BE REPLACED. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
318076 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAH

Patients

Seq Age Sex Outcome Treatment
1 21 YR Hospitalization