PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2014-03494
- Event Type
- Injury
- Date Received
- May 30, 2014
- Date of Event
- March 25, 2014
- Report Date
- May 15, 2014
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE REPORTED DEVICE OPERATES DIFFERENTLY THAN EXPECTED (NEEDLE TO CUFF MISS)/SUTURE RETRIEVAL ISSUE WAS CONFIRMED. BASED ON A VISUAL AND FUNCTIONAL INSPECTION ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. A QUERY OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. FIVE UNUSED, STERILE REPRESENTATIVE SAMPLES WITH THE LOT NUMBER 40107K1 WERE RETURNED FOR EVALUATION. FUNCTIONAL TESTING WAS PERFORMED AND THE DEVICES PASSED WITH ACCEPTABLE RESULTS. NO MALFUNCTION OR ABNORMAL OBSERVATIONS WERE DETECTED, BASED ON THE INVESTIGATION FINDINGS FROM THE TESTED SAMPLES.
(B)(4). ESTIMATED DATE OF OCCURRENCE, EXACT DATE WAS NOT REPORTED. (B)(4) - FAILURE TO FOLLOW STEPS/INSTRUCTIONS. PERFORM A FEMORAL ANGIOGRAM THROUGH THE INTRODUCER SHEATH TO VERIFY THAT THE ACCESS SITE IS IN THE COMMON FEMORAL ARTERY. TO AVOID POSTERIOR WALL SUTURE PLACEMENT AND POSSIBLE LIGATION OF THE ANTERIOR AND POSTERIOR WALLS OF THE FEMORAL ARTERY, FLUOROSCOPICALLY EVALUATE THE FEMORAL ARTERY FOR SIZE, CALCIUM, TORTUOSITY, AND FOR DISEASE OR DISSECTIONS OF THE ARTERIAL WALL. THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY WAS ATTEMPTED USING THE PROGLIDE DEVICE WITH A 5F SHEATH AFTER AN PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) INTERVENTIONAL PROCEDURE. REPORTEDLY, AFTER NEEDLE DEPLOYMENT THE PLUNGER WAS RETRACTED, AN ANTERIOR CUFF MISS OCCURRED. THE DEVICE WAS REMOVED AND HEMOSTASIS WAS ACHIEVED USING A SECOND PROGLIDE DEVICE. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE. NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE WAS REPORTED. NO FEMORAL ANGIOGRAM OR OTHER IMAGING WAS TAKEN. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 318623 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 40107K1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | SHEATH: 5F, HEPARIN |