FDA Adverse Event Malfunction Summary report: N

COLORADO NEEDLE, 3" STRAIGHT SLEEVE, 3MM 45 ANG

MDR report key: 3841530 · Received May 30, 2014

Report

Report Number
0008010177-2014-00142
Event Type
Malfunction
Date Received
May 30, 2014
Date of Event
May 8, 2014
Report Date
May 8, 2014
Manufacturer
STRYKER ORTHOPAEDICS-CORK
Product Code
GEI
PMA / PMN Number
K033232
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE VISUAL INVESTIGATION OF THE PICTURE PROVIDED CONFIRMED THAT THE TIP OF THE NEEDLE IS MISSING/BROKEN OFF. BECAUSE THE COMPLAINED DEVICE WAS NOT RETURNED IT WAS NOT POSSIBLE TO PERFORM FURTHER INVESTIGATIONS. THE REVIEW OF THE RELATED MANUFACTURING (BY THE SUPPLIER) AND INSPECTION DOCUMENTS (BY (B)(4)) SHOWED THAT THE LOT WAS MANUFACTURED ACCORDING TO THE SPECIFICATION. THE SUPPLIER TEAM HAS COMPLETED A THOROUGH INVESTIGATION OF ALL THE ASSOCIATED PROCESSING STEPS AND THE RELATED MATERIALS (INPUTS AND OUTPUTS). BASED ON THE AVAILABLE INFORMATION, THE ROOT CAUSE FOR THE REPORTED EVENT COULD NOT BE DETERMINED. AS PER STATISTICAL EVALUATION THERE WERE NO INDICATIONS FOR ANY SYSTEMATIC, DESIGN, MATERIAL OR MANUFACTURING RELATED ISSUE.

Additional Manufacturer Narrative · 1

DEVICE NOT AVAILABLE FOR EVALUATION. IF ADDITIONAL INFORMATION IS RECEIVED IT WILL BE REPORTED ON A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILST IN SURGERY, THE TIP OF A COLORADO NEEDLE HAS BROKEN OFF. THE TIP COULD NOT BE FOUND AND THE PATIENT REQUIRED AN X-RAY TO SEE IF IT HAD FALLEN IN THE WOUND. THE X-RAY WAS INCONCLUSIVE AND THEY ARE STILL UNAWARE OF WHERE THE TIP OF THE NEEDLE IS. THIS RESULTED IN A DELAY TO SURGERY.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILST IN SURGERY, THE TIP OF A COLORADO NEEDLE HAS BROKEN OFF. THE TIP COULD NOT BE FOUND AND THE PATIENT REQUIRED AN X-RAY TO SEE IF IT HAD FALLEN IN THE WOUND. THE X-RAY WAS INCONCLUSIVE AND THEY ARE STILL UNAWARE OF WHERE THE TIP OF THE NEEDLE IS. THIS RESULTED IN A DELAY TO SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
318380 COLORADO NEEDLE, 3" STRAIGHT SLEEVE, 3MM 45 ANG INSTRUMENT GEI STRYKER ORTHOPAEDICS-CORK 13275017

Patients

Seq Age Sex Outcome Treatment
1