FDA Adverse Event Malfunction Summary report: N

SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

MDR report key: 3841362 · Received May 30, 2014

Report

Report Number
1416980-2014-17452
Event Type
Malfunction
Date Received
May 30, 2014
Date of Event
May 6, 2014
Report Date
May 8, 2014
Manufacturer
BAXTER HEALTHCARE - SWINFORD
Product Code
KDJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED FOR EVALUATION. VISUAL INSPECTION OF THE RETURNED DEVICE IDENTIFIED A SMALL BLACK PARTICLE ON THE DISCONNECT CAP. THE SIZE OF THE PARTICLE WAS FOUND TO BE SMALL ENOUGH THAT IT WOULD NOT BE CONSIDERED A DEFECT PER THE PRODUCT'S SPECIFICATIONS. THE SUB ASSEMBLY LOT NUMBER OF THE DEVICE'S CAP COMPONENT WAS IDENTIFIED UPON RECEIPT OF THE DEVICE. A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS OF THE CAP. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. THE CAUSE OF THE PARTICULATE MATTER COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

A CUSTOMER REPORTED OBSERVING PARTICULATE MATTER IN THE CAP OF A UV FLASH PREP KIT. THIS WAS OBSERVED OUTSIDE OF PATIENT USE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
319486 SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - SWINFORD

Patients

Seq Age Sex Outcome Treatment
1