LIGACLIP** ENDOSCOPIC CLIP APPLIER
Report
- Report Number
- 3005075853-2014-03681
- Event Type
- Malfunction
- Date Received
- May 30, 2014
- Report Date
- April 1, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). RESULTS: YIELDED JAW. THE ANALYSIS RESULT SHOWED THAT THE ER320 INSTRUMENT WAS RECEIVED EMPTY AND THE JAWS WERE FOUND TO BE IN A YIELDED CONDITION MAKING THE DEVICE NON-FUNCTIONAL. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. NO FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE RETURNED CONDITION OF THE DEVICE. IT IS KNOWN FROM THE HISTORY OF THE DEVICE THAT THE CONDITION OF THE JAWS MAY LEAD DROPPING/EJECTED CLIPS. IN ADDITION, THE LOCKOUT FEATURE WAS NOTED TO BE NON FUNCTIONAL. IN ORDER TO EVALUATE THE CONDITION OF THE INTERNAL COMPONENTS OF THE DEVICE, IT WAS DISASSEMBLED. UPON DISASSEMBLING, THE LOCKOUT POSTS WERE FOUND TO BE BROKEN WHICH SUGGEST THAT A LOCKOUT PREMATURE OCCURRED. IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THIS FINDING. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
THE SALES REP REPORTED THAT DURING AN UNKNOWN PROCEDURE THE DEVICE JAMMED, WOULD NOT FIRE. IT IS NOT KNOWN HOW THE CASE WAS COMPLETED. NO PATIENT CONSEQUENCES REPORTED. THERE WAS NO ADDITIONAL INFORMATION. ONE DEVICE WILL BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 319343 | LIGACLIP** ENDOSCOPIC CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |