FDA Adverse Event Malfunction Summary report: N

GREENLIGHT MOXY FIBER OPTIC

MDR report key: 3841345 · Received May 30, 2014

Report

Report Number
2937094-2014-00413
Event Type
Malfunction
Date Received
May 30, 2014
Date of Event
May 20, 2014
Report Date
May 20, 2014
Manufacturer
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
Product Code
GEX
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER (B)(4): THE FIBER SHOWS A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE AT THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE CAN ROTATE INDEPENDENTLY OF METAL CAP; THE GLASS CAP EXHIBITS MODERATE DEVITRIFICATION AT THE OUTPUT WINDOW; THE METAL CAP EXHIBITS MODERATE CHAR. BASED ON THE ANALYSIS ABOVE, THE POTENTIAL FOR FORWARD FIRING MAY EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH WOULD LIMIT THE PERFORMANCE OF THE FIBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE AIMING BEAM FIRES STRAIGHT OUT OF THE FIBER @ 141,352 JOULES (FORWARD FIRING). WHILE WORKING ON A MEDIUM SIZED GLAND., A LOT OF PROSTATIC STONES WERE UNCOVERED IN TISSUE. DOCTOR WORKED AROUND STONES FIBER WOULD FLASH. AFTER A FEW MOMENTS LASER ENERGY BEGAN TO FIRE STRAIGHT OUT OF TIP. THE PROCEDURE WAS COMPLETED USING A SECOND FIBER. PATIENT OUTCOME WAS REPORTED AS "OK."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
318813 GREENLIGHT MOXY FIBER OPTIC POWERED LASER SURGICAL INSTRUMENT GEX AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) 0010-2400 351A

Patients

Seq Age Sex Outcome Treatment
1