DUROM ACETABULAR COMPONENT 50/44 CODE J
Report
- Report Number
- 9613350-2014-03513
- Event Type
- Injury
- Date Received
- May 13, 2014
- Date of Event
- October 25, 2010
- Report Date
- April 14, 2014
- Manufacturer
- ZIMMER GMBH
- Product Code
- KWA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
Narratives
THE MANUFACTURER DID NOT RECEIVE OR X-RAYS FOR REVIEW. AS NO LOT NUMBERS WERE PROVIDED FOR THE DEVICES, THE DEVICE HISTORY RECORDS COULD NOT BE REVIEWED. BASED ON AN EXTENSIVE INVESTIGATION OF EVENTS REPORTED FROM SEVERAL USER FACILITIES OUTSIDE THE USA, ZIMMER IDENTIFIED THAT THE MOST PROBABLY CAUSE FOR THE OUTCOME OBSERVED WAS A LOOSE OR UNSTABLE CUP THAT RESULTED FROM USE OF SURGICAL TECHNIQUES NOT CONSISTENT WITH THE MANUFACTURER'S RECOMMENDATIONS. AS A CORRECTIVE ACTION, A RETRAINING PROGRAM FOR USERS OUTSIDE THE USA WAS INITIATED IN NOVEMBER 2009 AND REPORTED TO THE (B)(6) AS REQUIRED. THE DUROM CUP REPORTED IN THIS CASE IS NOT MARKETED IN THE USA. A SIMILAR CUP, COMPATIBLE WITH THE METASUL LDH FEMORAL HEAD, IS CLEARED IN THE USA AND A CORRECTIVE ACTION FOR THIS PRODUCT WAS REPORTED TO THE FDA IN JULY 2008 AS NOTIFICATION Z-2415/2426-2008. SINCE THIS CASE IS RELATED TO THE ISSUES FOR WHICH ZIMMER IMPLEMENTED A CORRECTIVE ACTION THERE WILL BE NO FURTHER INVESTIGATION. (B)(4).
A PRODUCT LIABILITY CLAIM WAS RAISED. IT WAS REPORTED THAT THE PATIENT WAS IMPLANTED A DUROM ACETABULAR COMPONENT 50/44 CODE J ON THE LEFT SIDE ON (B)(6) 2008. THE PATIENT WAS REVISED ON (B)(6) 2011 DUE TO PAIN AND LOOSENING. ON (B)(6) 2010 THE PATIENT WAS DIAGNOSED WITH SIGNIFICANT PROBLEMS WITH PAIN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 285509 | DUROM ACETABULAR COMPONENT 50/44 CODE J | DUROM ACETABULAR COMPONENT AND METASUL | KWA | ZIMMER GMBH | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR | Hospitalization| R |