FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 3841219 · Received February 27, 2014

Report

Report Number
9615739-2014-00074
Event Type
Malfunction
Date Received
February 27, 2014
Date of Event
January 30, 2014
Report Date
January 30, 2014
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THAT THE BRAKE CASTERS HAS DAMAGED STEMS WHICH ALSO DAMAGED THE BRAKE CASTER SUPPORTS. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2007. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. AN INVESTIGATION WAS PERFORMED AND THIS ISSUE WAS ADDRESSED UNDER AN ENGINEERING CHANGE. THE TECH REPLACED THE BRAKE CASTERS ANS SUPPORTS TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE ACCOUNT REPORTED THE BRAKES ARE NOT HOLDING. THERE WAS NO PT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
121651 CAREASSIST BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment
1