HEARTWARE® VENTRICULAR ASSIST SYSTEM
Report
- Report Number
- 3007042319-2014-00554
- Event Type
- Malfunction
- Date Received
- May 30, 2014
- Date of Event
- September 25, 2013
- Report Date
- April 30, 2014
- Manufacturer
- HEARTWARE, INC
- Product Code
- DSQ
- PMA / PMN Number
- P100047
- Removal / Correction Number
- Z-1607-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
THE DEVICE HAS BEEN RECEIVED AND EVALUATION IS ONGOING. PRELIMINARY EVALUATION AND TESTING REVEALED THAT THE RETURNED BATTERY CONTAINED A FAULTY CELL. THIS FINDING WAS RE-ASSESSED PER SOP REQUIREMENTS, AND DETERMINED TO BE REPORTABLE. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN THIRTY (30) DAYS OF RECEIPT. EVALUATION IN PROGRESS.
THE BATTERY WAS RETURNED; VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE BATTERY IN RELATION TO THE REPORTED EVENT. THOROUGH EXTERNAL VISUAL INSPECTION OF THE BATTERY REVEALED NO SIGNS OF PHYSICAL DAMAGE OR CONTAMINATION. FUNCTIONAL TESTING REVEALED THAT THE RETURNED BATTERY CONTAINED A FAULTY CELL PAIR. AN INTERNAL INVESTIGATION (CAPA) HAS BEEN OPENED TO ADDRESS THE ISSUE.
APPROXIMATELY TWO YEARS AND TWO MONTHS POST HVAD IMPLANTATION, THE SITE REPORTED THAT THE PATIENT'S BATTERY WOULD NOT CHARGE WHEN CONNECTED TO THE BATTERY CHARGER. THE BATTERY WAS REMOVED FROM THE PATIENT AND A NEW BATTERY WAS SUPPLIED. NO HARM OR INJURY TO THE PATIENT WAS REPORTED AS A RESULT OF THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 318887 | HEARTWARE® VENTRICULAR ASSIST SYSTEM | CIRCULATORY ASSIST SYSTEM, BATTERY | DSQ | HEARTWARE, INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |