FDA Adverse Event Malfunction Summary report: N

AFFINITY THREE BIRTHING BED

MDR report key: 3841127 · Received February 26, 2014

Report

Report Number
1824206-2014-00651
Event Type
Malfunction
Date Received
February 26, 2014
Date of Event
January 30, 2014
Report Date
January 30, 2014
Manufacturer
HILL-ROM, INC.
Product Code
HDD
PMA / PMN Number
K915779
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE RIGHT SIDE RAIL DAMPENER WAS LOCKED IN PLACE AND NOT MOVING WHEN THE SIDE RAIL WAS LIFTED SO THE SIDE RAIL COULD NOT LIFT TO THE LATCH POSITION. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2013. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE RIGHT SIDE RAIL DAMPENER TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE ACCOUNT REPORTED THE RIGHT SIDE RAIL WOULD NOT LATCH. THE BED WAS LOCATED IN LABOR AND DELIVERY AT THE FACILITY. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
117717 AFFINITY THREE BIRTHING BED BIRTHING BED HDD HILL-ROM, INC. 3700

Patients

Seq Age Sex Outcome Treatment
1