HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3841099 · Received May 30, 2014

Report

Report Number
1416980-2014-17430
Event Type
Malfunction
Date Received
May 30, 2014
Date of Event
May 6, 2014
Report Date
May 6, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CAUSE FOR THE REPORTED CONDITION WAS DETERMINED TO BE A LOOSE CONNECTION BECAUSE A LOOSE CONNECTION IS A KNOWN CAUSE OF THIS TYPE OF ALARM. THE DEVICE WAS NOT RETURNED; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR (SE) 2240 (AIR IN SET) ALARM OCCURRED DURING DWELL THREE OF FIVE ON THE HOME CHOICE (HC). THE HOME PATIENT (HP) WAS CONNECTED AT THE TIME OF THE EVENT. THE HP INDICATED THAT THE CONNECTION TO A SUPPLY BAG WAS LOOSE. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HP IN CLEARING THE ALARM AND ENDING THERAPY. THE TSR ADVISED THE HP TO START OVER WITH NEW SUPPLIES. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.