XMAX MOTOR
Report
- Report Number
- 1045834-2014-12100
- Event Type
- Malfunction
- Date Received
- May 30, 2014
- Date of Event
- May 1, 2014
- Report Date
- May 13, 2014
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- ERL
- PMA / PMN Number
- PK965080
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THERE WAS A HOLE IN THE HOSE. THEREFORE, THE REPORTED CONDITION OF LEAKING WAS CONFIRMED. THE HEAT TEST COULD NOT BE COMPLETED DUE TO THE HOLE IN THE HOSE. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO NORMAL WEAR OF COMPONENT PARTS OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
ADDITIONAL NARRATIVE:THE ACTUAL DEVICE HAS BEEN RETURNED AND IS CURRENTLY PENDING EVALUATION. ONCE RELIABILITY ENGINEERING EVALUATES THE DEVICE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT PRIOR TO SURGERY IT WAS DISCOVERED THAT THE MOTOR DEVICE WAS ¿GETTING HOT¿ AND MIGHT HAVE BEEN LEAKING OIL. THERE WERE NO DELAYS TO THE PLANNED SURGICAL PROCEDURE. A SPARE IDENTICAL DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF EVENT WAS UNKNOWN ALTHOUGH THE REPORTER CLARIFIED THE EVENT OCCURRED ON (B)(6) 2014 . ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 319461 | XMAX MOTOR | DRILL, SURGICAL, ENT (PNEUMATIC) - HANDPIECE | ERL | DEPUY SYNTHES POWER TOOLS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |