FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 3841088
·
Received May 30, 2014
Report
- Report Number
- 1416980-2014-17429
- Event Type
- Malfunction
- Date Received
- May 30, 2014
- Date of Event
- May 7, 2014
- Report Date
- May 7, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT. THE DEVICE WAS VISUALLY INSPECTED, FUNCTIONALLY TESTED AND A POWER ON SELF-TEST PERFORMED. THE CAUSE OF THE DOOR OPEN ALARM WAS DETERMINED TO BE DUE TO A MISSING DOOR MAGNET. TO CORRECT THE CONDITION, THE MAGNET WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
DURING PRODUCT SERVICE BY A SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE A DOOR OPEN ALARM. NO ADDITIONAL INFORMATION IS AVAILABLE.