FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3841088 · Received May 30, 2014

Report

Report Number
1416980-2014-17429
Event Type
Malfunction
Date Received
May 30, 2014
Date of Event
May 7, 2014
Report Date
May 7, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT. THE DEVICE WAS VISUALLY INSPECTED, FUNCTIONALLY TESTED AND A POWER ON SELF-TEST PERFORMED. THE CAUSE OF THE DOOR OPEN ALARM WAS DETERMINED TO BE DUE TO A MISSING DOOR MAGNET. TO CORRECT THE CONDITION, THE MAGNET WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING PRODUCT SERVICE BY A SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE A DOOR OPEN ALARM. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
319239 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1