FDA Adverse Event Malfunction Summary report: N

ECHELON*FLEX60 LINEAR CUTTER

MDR report key: 3840903 · Received May 30, 2014

Report

Report Number
3005075853-2014-03674
Event Type
Malfunction
Date Received
May 30, 2014
Date of Event
May 7, 2014
Report Date
May 12, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K081146
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AT THE TIME OF THIS SUBMISSION, THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS. IF THE DEVICE IS RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE ANALYSIS FOUND THAT ONE EC60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH ONE ECR60B CARTRIDGE LOADED IN THE DEVICE. THE CARTRIDGE RELOAD WAS RECEIVED FULLY FIRED AND IN GOOD VISUAL CONDITION. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. IT SHOULD BE NOTED THAT IN ORDER TO OPEN A DEVICE THAT HAS BEEN PARTIALLY FIRED OR FULLY FIRED A REVERSE STROKE NEEDS TO BE PERFORMED TRIGGER TO TRIGGER TO HANDLE IN ORDER TO RETURN THE KNIFE TO THE HOME POSITION (INDICATOR IN THE "0" POSITION) AND PRESS THE ANVIL RELEASE BUTTON TO OPEN. EVENT COULD NOT BE CONFIRMED AS THE DEVICE CLOSED AND OPENED WITHOUT ANY DIFFICULTIES NOTED. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC-ASSISTED DISTAL GASTRECTOMY (LADG), THE JAWS DID NOT OPEN AFTER THE FIRING. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
318140 ECHELON*FLEX60 LINEAR CUTTER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA L4E239

Patients

Seq Age Sex Outcome Treatment
1 ECR60B(LOT#:L4E67E)X1