FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3840880 · Received May 30, 2014

Report

Report Number
1416980-2014-17418
Event Type
Malfunction
Date Received
May 30, 2014
Date of Event
May 7, 2014
Report Date
May 7, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN, A DEVICE ANALYSIS COULD NOT BE COMPLETED. HOWEVER, THE HOME PATIENT (HP) REPORTED DISCONNECTING THE HEATER BAG AND REPLACING IT WITH A SUPPLY BAG FROM A SEPARATE CASSETTE LINE. USE ERROR AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿ WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE WARNS THE USER NOT TO REPLACE EMPTY SOLUTION BAGS OR RECONNECT DISCONNECTED SOLUTION BAGS DURING THERAPY. IT ALSO WARNS THE USER THAT POSSIBLE CONTAMINATION OF THE FLUID OR FLUID PATHWAYS CAN RESULT IF DISPOSABLES ARE REUSED. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

WHILE TROUBLESHOOTING FOR A CHECK LINES AND BAGS ALARM, A TECHNICAL SERVICE REPRESENTATIVE (TSR) WAS INFORMED THAT A CARE GIVER (CG) DISCONNECTED THE HEATER BAG AND REPLACED IT WITH ONE OF THE USED SUPPLY BAGS. THE SWAP OCCURRED DURING THE FILL STAGE OF AUTOMATED PERITONEAL DIALYSIS (APD) USING A HOMECHOICE (HC) DEVICE. THE HOME PATIENT (HP) WAS CONNECTED AT THE TIME OF THE DISCONNECTION AND RECONNECTION OF THE BAGS. THE TSR REVIEWED PROPER PROCEDURES WITH THE CG AND INSTRUCTED THE CG TO CONTACT A REGISTERED NURSE (RN) REGARDING THE POTENTIAL CONTAMINATION. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
318533 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 67 YR HOMECHOICE