FENESTRATED BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-03342
- Event Type
- Malfunction
- Date Received
- May 30, 2014
- Date of Event
- March 26, 2014
- Report Date
- May 27, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION FOUND THE INSTRUMENT'S PITCH CABLE WAS BROKEN. THE PITCH CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINED THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. FAILURE ANALYSIS INVESTIGATION ALSO FOUND THE FOLLOWING: THE INSTRUMENT'S DISTAL PULLEY HAD SCRATCH MARKS AND ABRASIONS. THERE WERE SCRATCHES ON THE SURFACE OF THE DISTAL PULLEY. THE INSTRUMENT'S GRIP TIPS WERE BENT. ONE GRIP IS BENT, CAUSING SIDE TO SIDE MISALIGNMENT OF THE GRIPS. THERE IS A .075 OFFSET AT THE TIPS, INDICATING OVERLOADING AT THE TIP. THE INSTRUMENT PASSED ELECTRICAL CONTINUITY TESTING. FAILURE ANALYSIS CONCLUDED THAT THE BENT DAMAGE WAS LIKELY DUE TO MISHANDLING/MISUSE. NO OTHER DAMAGE WAS FOUND. THE ENDOWRIST® INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT IN ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE BROKEN CABLE IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING A DA VINCI PARAAORTIC LND SURGICAL PROCEDURE, IT WAS NOTED THAT THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WOULD NOT FOLLOW THE ORDERS OF THE SURGEON. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED WITH ANOTHER INSTRUMENT. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 318226 | FENESTRATED BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420205-05 | M10131104 113 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |