GREENLIGHT MOXY FIBER OPTIC
Report
- Report Number
- 2937094-2014-00424
- Event Type
- Malfunction
- Date Received
- May 30, 2014
- Date of Event
- April 1, 2014
- Report Date
- April 1, 2014
- Manufacturer
- AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
- Product Code
- GEX
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
FAILURE ANALYSIS FOR FIBER (B)(4): THE CONNECTOR APPEARS IN GOOD CONDITION; THE CONNECTOR WAS ABLE TO BE CONNECTED TO THE DEVICE COUPLER BOARD; NO FAILURE WAS FOUND WITH THE CONNECTOR; THE FIBER SHOWS A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE AT THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE CAN ROTATE INDEPENDENTLY OF METAL CAP; THE METAL CAP EXHIBITS MODERATE CHAR AND DETRITUS BUILDUP; THE GLASS CAP EXHIBITS SEVERE DEVITRIFICATION AT THE OUTPUT WINDOW; THE OUTER FLOW TUBING EXHIBITS SEVERE CONTAMINATION, LIKELY BIOLOGIC. BASED ON THE ANALYSIS ABOVE, THE POTENTIAL FOR FORWARD FIRING MAY EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PRODUCT COMPLAINT OF "CONNECTION ERROR" COULD NOT BE CONFIRMED; A CAP FRACTURE WAS OBSERVED; THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH WOULD LIMIT THE PERFORMANCE OF THE FIBER.
IT WAS REPORTED THAT WHILE USING THE SURGICAL FIBER DURING A PROSTATE PROCEDURE, THERE WAS A "CONNECTION ERROR, UNABLE TO ROTATE FIBER" AT 3 MINUTES AND 01 SECONDS AND 13,075 JOULES OF USE. THE PROCEDURE WAS COMPLETED USING A SECOND FIBER. PATIENT OUTCOME: "NO DAMAGES TO THE PATIENT".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 318223 | GREENLIGHT MOXY FIBER OPTIC | POWERED LASER SURGICAL INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) | 0010-2400 | 342A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |