FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 3840536 · Received May 30, 2014

Report

Report Number
1416980-2014-17367
Event Type
Malfunction
Date Received
May 30, 2014
Date of Event
December 8, 2013
Report Date
May 7, 2014
Manufacturer
BAXTER HEALTHCARE
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAD A SYSTEM ERROR 2240 (AIR IN LINE) ON THE HOMECHOICE (HC) DEVICE DURING AN UNSPECIFIED STEP OF PERITONEAL DIALYSIS (PD) THERAPY. PATIENT CONNECTION AT THE TIME OF THE ALARM IS UNKNOWN. THE REPORTER STATED THAT IT WAS IMPOSSIBLE TO CONTINUE PD THERAPY AFTER THE ALARM. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
318809 SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1 HOMECHOICE PRO