FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 3840529 · Received May 30, 2014

Report

Report Number
2024168-2014-03450
Event Type
Injury
Date Received
May 30, 2014
Date of Event
March 26, 2014
Report Date
May 6, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. BECAUSE KEY COMPONENTS WERE NOT RETURNED WITH THE DEVICE, THE SCOPE OF THIS INVESTIGATION WAS VERY LIMITED AND THE REPORTED CUFF MISS COULD NOT BE CONFIRMED. BASED ON ANALYSIS OF THE RETURNED DEVICE, NO QUALITY DEFICIENCY WAS DETECTED. A SEARCH OF THE LOT HISTORY RECORD DID NOT REVEAL ANY NON-CONFORMANCES ASSOCIATED WITH THIS LOT WHICH COULD HAVE CONTRIBUTED TO THIS COMPLAINT AND ITS INVESTIGATION FINDINGS. A REVIEW OF THE COMPLAINT DATABASE FOR THIS LOT DID NOT REVEAL ANY INDICATION OF A LOT SPECIFIC PRODUCT QUALITY DEFICIENCY. BASED ON THE INVESTIGATION, THERE DOES NOT APPEAR TO BE ANY INDICATION OF A PRODUCT QUALITY DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE AFTER AN UNSPECIFIED PROCEDURE. REPORTEDLY, A CUFF MISS OCCURRED. A SECOND PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE PROGLIDE DEVICE. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE ENTIRE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
319500 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 31126K1

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention