HOMECHOICE
Report
- Report Number
- 1416980-2014-17372
- Event Type
- Malfunction
- Date Received
- May 30, 2014
- Date of Event
- May 6, 2014
- Report Date
- May 6, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE HOMECHOICE WAS RECEIVED, AND A DEVICE ANALYSIS IS CURRENTLY UNDERWAY. UPON COMPLETION OF THE EVALUATION OR IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). THE HOMECHOICE (HC) DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE EVENT HISTORY LOG REVIEW SHOWED NO KEYSTROKES, PROGRAMMING, OR USE RELATED EVENTS THAT INDICATED AND/OR CONTRIBUTED TO THE REPORTED ISSUE. THE EXTERNAL/INTERNAL INSPECTION WAS PERFORMED WITH NO ISSUES NOTED. TEMPERATURE WAS WITHIN SPECIFICATIONS. THE DEVICE PASSED ELECTRICAL TESTING. A VOLUMETRIC ACCURACY TEST WAS PERFORMED AND THE DEVICE FAILED. TEST ARTICLE, PISTON FOAM, WAS INSTALLED AND THE DEVICE PASSED THE VOLUMETRIC ACCURACY TEST. THE EVALUATION REVEALED THE CAUSE OF THE FAILURE TO BE DETERIORATED PISTON FOAM. THE PISTON FOAM WAS TO BE SCRAPPED AND THE DEVICE WAS SENT FOR SERVICING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE FAILED THERAPY MONITORED FLUID VOLUME TESTING. THERE WAS NO PATIENT INVOLVED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 319497 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |