SIMPLEX TOBRAMYCIN 1 PK
Report
- Report Number
- 0002249697-2014-01975
- Event Type
- Malfunction
- Date Received
- May 30, 2014
- Date of Event
- May 5, 2014
- Report Date
- May 5, 2014
- Manufacturer
- STRYKER ORTHOPAEDICS-MAHWAH
- Product Code
- LOD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- OTHER
Narratives
AN EVENT REGARDING SETTING TIME INVOLVING SIMPLEX BONE CEMENT WAS REPORTED. THE EVENT WAS NOT CONFIRMED. VISUAL INSPECTION AND FUNCTIONAL TESTING WAS COMPLETED ON THREE RETURNED SAMPLES OF THE REPORTED LOT. THE RESULTS WERE SATISFACTORY AND WITHIN SPECIFICATION. BMR REVIEW FOR THE SPECIFIED LOT INDICATES THAT THE DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. CHR REVIEW CONFIRMED ONE OTHER SIMILAR EVENT FOR THE REPORTED LOT. THE INVESTIGATION CONCLUDED THAT THE REPORTED SETTING TIME ISSUE CANNOT BE CONFIRMED. THE MIXING PROPERTIES OF THE RETURNED SAMPLES OF THE REPORTED LOT WERE TESTED AND SHOW THAT ALL REQUIRED SPECIFICATIONS ARE MET. NO FURTHER INVESTIGATION FOR THIS EVENT IS POSSIBLE AT THIS TIME.
ADDITIONAL INFORMATION HAS BEEN REQUESTED AND IF RECEIVED, WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT. THE SUBJECT DEVICE IS NOT CLEARED FOR SALE IN THE U.S., BUT A SIMILAR DEVICE IS COMMERCIALLY AVAILABLE IN THE U.S.
IT WAS REPORTED BY THE SALES REP THAT THE CEMENT WAS HARDENING TO SLOW. THE HOSPITAL EXPERIENCED THAT THE CEMENT TOOK VERY LONG TIME TO HARDEN AND EVEN THOUGH THE MIXING WAS DONE THE RIGHT WAY THE CEMENT SEEM UNEVEN/UNEQUAL. PATIENT NR 2: THE CEMENT WAS 1 PORTION FROM THE CUPBOARD AT THE OR WITH LOT TLU062 . KEPT IN 21 DEGREES C + 1 PORTION (ANOTHER LOT) FROM THE STORAGE ROOM WHICH IS KEPT IN 21 DEGREES C. MIXED AS USUAL... HARDING TIME 17 MINUTES. CEMENT SEEM UNEVEN/UNEQUAL.
IT WAS REPORTED BY THE SALES REP THAT THE CEMENT WAS HARDENING TO SLOW. THE HOSPITAL EXPERIENCED THAT THE CEMENT TOOK VERY LONG TIME TO HARDEN AND EVEN THOUGH THE MIXING WAS DONE THE RIGHT WAY THE CEMENT SEEM UNEVEN/UNEQUAL. PATIENT NR 2: THE CEMENT WAS 1 PORTION FROM THE CUPBOARD AT THE OR WITH LOT TLU062 . KEPT IN 21 DEGREES C + 1 PORTION (ANOTHER LOT) FROM THE STORAGE ROOM WHICH IS KEPT IN 21 DEGREES C. MIXED AS USUAL¿HARDING TIME 17 MINUTES. CEMENT SEEM UNEVEN/UNEQUAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 319477 | SIMPLEX TOBRAMYCIN 1 PK | IMPLANT | LOD | STRYKER ORTHOPAEDICS-MAHWAH | TLU062 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |