ANALYTICAL E MODULE
Report
- Report Number
- 1823260-2014-03871
- Event Type
- Malfunction
- Date Received
- May 30, 2014
- Date of Event
- May 14, 2014
- Report Date
- June 25, 2014
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K961481
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A ROOT CAUSE COULD NOT BE DETERMINED WITH THE INFORMATION PROVIDED. ADDITIONAL DATA WERE REQUESTED BUT NOT PROVIDED. THE PROVIDED QUALITY CONTROL RESULTS WERE WITHIN SPECIFICATION. IT WAS NOTED THE CUSTOMER WAS NOT PERFORMING QUALITY CONTROL ON EACH REAGENT PACK SEPARATELY AS RECOMMENDED.
THIS EVENT OCCURRED IN (B)(6).
THE CUSTOMER PROVIDED THE FOLLOWING INFORMATION. THE RESULT OF 2134 MIU/ML WAS NOT AN ACTUAL PATIENT RESULT.
THE CUSTOMER ALLEGED THEY RECEIVED A QUESTIONABLE INTACT HUMAN CHORIONIC GONADOTROPIN + THE SS-SUBUNIT (HCGB) RESULT FOR ONE PATIENT ON THEIR E- MODULE. THE PATIENT'S INITIAL HCGB RESULT WAS 0.100 MIU/ML ACCOMPANIED BY A DATA FLAG AND IT WAS REPORTED OUTSIDE THE LABORATORY. THE PATIENT DOUBTED THE RESULT. ON (B)(6) 2014, THE SAMPLE WAS REPEATED ON THE ORIGINAL E-MODULE. THE RESULT WAS 2131 MIU/ML ACCOMPANIED BY A DATA FLAG. THE CUSTOMER ALSO PROVIDED THE RESULT 2134 MIU/ML. CLARIFICATION ON WHAT THIS RESULT WAS HAS BEEN REQUESTED. ON (B)(6) 2014, THE SAMPLE WAS TESTED ON ANOTHER E-MODULE AND THE RESULT WAS 2146 MIU/ML ACCOMPANIED BY A DATA FLAG. THERE WERE NO DEATHS, INJURIES, ILLNESSES, OR DETERIORATIONS IN HEALTH ASSOCIATED WITH THE ERRONEOUS RESULTS. INFORMATION ON WHETHER THE PATIENT WAS HARMED BY ANY ACTION TAKEN WAS REQUESTED BUT NOT PROVIDED. THE HCGB REAGENT LOT NUMBER WAS 173268. THE EXPIRATION DATE WAS REQUESTED BUT NOT PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 318662 | ANALYTICAL E MODULE | IMMUNOCHEMISTRY ANALYZER | JJE | ROCHE DIAGNOSTICS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 034 YR |