COULTER LH 750 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2014-01131
- Event Type
- Malfunction
- Date Received
- May 30, 2014
- Date of Event
- April 30, 2014
- Report Date
- April 30, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K011342
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT. THE FSE FOUND A LEAK FROM COMPLETE BLOOD COUNT (CBC) LYSE RESTRICTOR TUBING DUE TO SMALL PIN HOLE AT FITTING SIDE OF THE TUBING. THE FSE REPLACED TUBING AND PERFORMED CBC LYSE PRIME AND VERIFIED NO FURTHER LEAKING. THE INSTRUMENT WAS RETURNED TO NORMAL OPERATION. (B)(4).
THE CUSTOMER REPORTED INTERNAL/EXTERNAL LEAKING AT THE RIGHT SIDE OF THE COULTER LH 750 HEMATOLOGY ANALYZER WHILE RUNNING SAMPLES. THE VOLUME OF THE LEAK WAS APPROXIMATELY 5 ML OF CLEAR FLUID. THE LEAK WAS CONTAINED WITHIN THE INSTRUMENT. THERE WAS NO REPORTED EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. THE CUSTOMER WAS WEARING GLOVES AND LAB COAT AT THE TIME OF THE INCIDENT. NO ERRONEOUS RESULTS WERE REPORTED OUT OF THE LAB. THERE WAS NO DEATH, INJURY, OR AFFECT TO USER OR TO PATIENT TREATMENT ATTRIBUTED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 319070 | COULTER LH 750 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |