FENESTRATED BIPOLAR FORCEPSINSTRUMENT
Report
- Report Number
- 2955842-2014-03325
- Event Type
- Malfunction
- Date Received
- May 29, 2014
- Date of Event
- April 28, 2014
- Report Date
- May 2, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS CONFIRMED THE REPORTED COMPLAINT. THE PITCH DOWN CABLE WAS FRAYED AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. OTHER CABLES AT THE WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE FRAYED CABLE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT DURING A DA VINCI TRIFLAP PROCEDURE THAT WAS STARTED AROUND 2 P.M AND COMPLETED ABOUT 6 P.M. DURING THE PROCEDURE, A SURGEON SAID SECOND ARM MOVEMENT IS NOT GOOD. IT DOESN'T SEEM TO HAVE ANY ENERGY AND A NURSE CHECKED AND FOUND THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WIRE HAS BEEN BROKEN. BECAUSE THIS EVENT WAS OCCURRED AT THE END OF THE PROCEDURE, THE CUSTOMER COMPLETED IT WITHOUT CHANGING INSTRUMENT. THE SURGEON IS EXPERIENCED AND HE COMPLAINED WHY WIRES HAVE BEEN BROKEN TOO MUCH WHEN HE PERFORMS SURGERIES AND SAID THAT HE WILL SAY THIS ISSUE TO INTUITIVE SURGICAL DIRECTOR. NOTHING REPORTEDLY FELL INTO A PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 317008 | FENESTRATED BIPOLAR FORCEPSINSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420205-05 | M11131130 510 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR |