FDA Adverse Event Malfunction Summary report: N

MULTI-LINK RX ZETA CORONARY STENT SYSTEM

MDR report key: 3836623 · Received May 28, 2014

Report

Report Number
2024168-2014-03365
Event Type
Malfunction
Date Received
May 28, 2014
Date of Event
May 6, 2014
Report Date
May 7, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
MAF
PMA / PMN Number
P020047
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE AND PACKAGING WERE RETURNED FOR EVALUATION. THE PRODUCT MIX-UP WAS CONFIRMED. IN THIS CASE, THERE IS NO INDICATION THAT THE MIX-UP OCCURRED DURING MANUFACTURING OF THE PRODUCTS AT ABBOTT VASCULAR. IT IS POSSIBLE THAT THE 3.0X38MM ZETA POUCH/DEVICE WAS INADVERTENTLY PLACED IN A 3.5X38MM RX ZETA CHIPBOARD BOX AT THE ACCOUNT, ALTHOUGH THIS CANNOT BE CONFIRMED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. A QUERY OF THE COMPLAINT HANDLING DATABASE REVEALED NO OTHER SIMILAR INCIDENTS REPORTED FROM THIS LOT. ALTHOUGH THE EXACT CAUSE FOR THE MISLABELING/MIX-UP CANNOT BE DETERMINED, THERE IS NO INDICATION OF A MANUFACTURING OR PRODUCT QUALITY DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT A ZETA STENT DELIVERY SYSTEM WAS BEING UNPACKED FOR USE; HOWEVER, THE CHIPBOARD BOX WAS LABELED 3.5 X 38 MM, LOT NUMBER 3100441, EXPIRATION 9/2015 AND THE POUCH WAS LABELED 3.0 X 38 MM, LOT NUMBER 3042641, EXPIRATION 3/2015. THE DEVICE WAS LABELED 3.0 X 38, LOT # 3042641. THE DEVICE WAS NOT USED. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
312571 MULTI-LINK RX ZETA CORONARY STENT SYSTEM CORONARY STENT SYSTEM MAF AV-TEMECULA-CT 3100441

Patients

Seq Age Sex Outcome Treatment
1